Installation Data - Kodak 104 Service Manual

Medical x-ray processor

Advertisement

Installation Data

Installation Data
77 cm (30.3")
Optional working ta-
ble or closed base
cabinet (accessory)
75 cm (29.5")
75 cm (29.5")
1.
Wall socket
120 V, 15A (depending on
machine model). Power lead
should be equipped with GFI,
Earth-Leakage Switch, 25 A /
30 mA nominal error-current.
In addition, a wall disconnect
switch can be installed.
2.
Filtered fresh water connection
19 mm (3/4 in.) with stop cock,
permissible pressure 20 - 85
psi, water temperature
5 - 25°C.
3.
Drainage plastic pipe (PVC)
∅ 50 mm (2 in.) incl. syphon.
4.
Drainage resp. collecting con-
tainers for used developer.
5.
Drainage resp. collecting con-
tainers for used fixer.
6.
Storing space for replenish-
ment tanks: Below machine or
externally.
7.
Ventilation of darkroom is nec-
essary. 10 room air exchanges
per hour.
8.
Sink with freshwater and flexi-
ble hose. Inner dimensions
minimum (LxWxH)
60x40x30 cm (24x16x12 in.).
59 cm (23.2")
59 cm (23.2")
59 cm (23.2")
Darkroom
8
Sink
60x40cm (24x16'')
Measures and positions are recommendations
7
50cm (20'')
1
2
4 5
6
3
3