Processor Description; Special Tools; Specifications And Data - Kodak M35 Service Manual

Processor
Hide thumbs Also See for M35:
Table of Contents

Advertisement

The KODAK M35 and M35A X-OMAT Processors
are designed to process medical x-ray film in either
sheets or rolls. This tabletop processor is
self-threading and has a 3-rack processing section
with a dryer.

Special Tools

The following special tools are necessary for some of
the service procedures in this manual.
• TL-3230
Sealant
• TL-2244
Motor Oil, 1-oz Tube
• TL-2324
Motor Oil, 12-oz Tube
• TL-1481
Potentiometer Adjusting Tool
• 761217
Thermometer
Film Sizes
• 10 - 35 cm (4 - 14 in.) width
• 10 cm (4 in.) minimum length
• 4.6 m (15 ft) maximum length
Processor Speed
approximately
76 cm (30 in.) per minute
Control of the Solution Temperature
The temperature of the developer is adjustable from
approximately 30 to 46°C (85 to 115°F). The
temperature of the fixer is not controlled by
adjustment.
1-2

Processor Description

Specifications and Data

Wash Water
The water for washing the film comes from the
customer's water supply. The FLOW CONTROL
VALVE regulates the usage at approximately 1 litre
1
( ⁄
gal) per minute.
4
Dryer
The processor has a roller transport dryer with
slotted AIR TUBES. A thermostat controls the
temperature from 46 - 65°C (115 - 150°F).
Power Requirements
The M35 Processor uses 200, 208, 220, or 240
volts with a single-phase, 2-wire plus earth ground or
220 or 240 volts with a 2-phase, 3-wire plus earth
ground and 50/60 Hz.
The M35A Processor uses 120 volts and a
single-phase, 2-wire plus an earth ground and
50/60 Hz.
H112_0089AC