Using the Analyzer's I/O Port

This chapter describes how to use the I/O port of the analyzer with the GPIB. For general
operation of the I/O port, see the
The I/O port on the analyzer's rear panel communicates with external devices (for example, a
handler on a production line).
The I/O port consists of the following 15 TTL compatible signals:
8-bit output
4-bit input
3 grounds
The signals IN 0 to IN 3 and OUT 0 to OUT 7 can be read and set by GPIB commands.
The Instrument BASIC commands
control the 8-bit I/O port without using GPIB commands. This operation is faster
than using an GPIB command. For more information on these commands, see the
Using Instrument BASIC with the 4396B
4396B Function Reference
Figure 10-1. I/O Port
manual.