Set List Table - Agilent Technologies 4396B Gpib Programming Manual

Network/spectrum/impedance analyzer
Hide thumbs Also See for 4396B:
Table of Contents

Advertisement

Line 110 de nes the table (array
center frequency, frequency span, number of points, power level, and resolution band width of
each segment as follows:
Segment
1
2
. .
.
Lines 130 to 150 call the subroutine
you de ned in line 120 (The analyzer can retain up to 31 segments).
d
The loop, lines 170 to 230, determines if you want to edit the table and con rms that the
segment is in the table.

Set List Table

In this portion of the program, the list table (edited using BASIC) is transferred to the analyzer.
What are other
commands are used to
set the list values?
Center
Frequency
Frequency
Span
. .
. .
.
.

When setting segment frequencies, instead of setting the
center/span de nition by using the
parameter
using:
1.
2.

When setting the IF band width (with the analyzer in
network analyzer mode), use the
) used to hold the list values. It contains the
Number of
Points
. .
.
(line 500) to edit and print as many segments as
Start to set the list table
Clear existing limit table
Start to set a segment
Enter center frequency
Enter frequency span
Enter number of points
Enter power level
Enter resolution band width
Complete the segment
Complete the list table
Activate list sweep
commands, you can de ne start/stop frequency by
/
parameter
parameter
/
commands (Maker to start/stop)
Power Level Resolution
Band Width
. .
. .
.
.
parameter
commands
command.
parameter
Using the List Sweep Function
a
/
9-3