Using the External RUN/CONT Connector

You can use the RUN or CONT commands in a program by inputting a TTL-compatible signal to
the External RUN/CONT connector on the rear panel. At the negative-going edge of a pulse
more than 20
s wide (T
) in the LOW state will trigger RUN or CONT.

p
Figure 7-2. RUN/CONT Trigger Signal

File System Exceptions

The analyzer supports both the LIF and DOS le formats. When using an LIF format disk, the
CREATE and CREATE DIR commands will generate an error.
Because the analyzer does not support an external disk drive, the MASS STORAGE IS
(MSI) statement cannot specify volumes other than the built-in disk drive (volume speci er
\INTERNAL,4", the default volume) and RAM disk memory (volume speci er \MEMORY,0").
Program I/O 7-3