Return to Execute GPIB command

Lines 230 to 270 clear SRQ before returning to the main routine.
Line 230 reads the analyzer's status byte. The
statement reads the Status
Byte Register of the address
(analyzer), and enters the value into . The command
error causes the SRQ and with bit 5 and bit 6 of the Status Byte Register set, the value of is
96. Reading the Status Byte Register by using the
command clears SRQ (status byte bit
6).
In line 240 and line 250, the command
reads the contents of the Standard Event Status
Register. With Bit 5 of Standard Event Status Register set, the value of
is 32. Reading
the Standard Event Status Register by using the
command clears the register.
A branch to
disables the interrupt. Therefore, the return from
must
reenable the interrupt.
Synchronizing the Analyzer from Remote
3-7