Lines 50 to 100 set the measurement coecient, Fadj, Fs, frequency span, resolution
bandwidth, number of measurement point, and averaging factor to typical values.
In lines 610 to 760, the subprogram
measurement points, in the area of center frequency,
total power is calculated by the following equation:
Where,
(
)
Power density spectrum [dBm/Hz]
D
x
Measurement frequency span [Hz]
F span
Number of measurement points
N op
I1, I2
The measurement point of left and right edge of a channel bandwidth(Fs)

Occupied Power Bandwidth Calculation

This program calculates the occupied bandwidth of the carrier signal. It rst computes the
combined power of all signal responses contained in the trace. It then calculates the point for
which 0.5 % of the total power lies to the right of the right maker and to the left of the left
marker.
Before running the program, set up the measurement, calibrate the analyzer, and connect the
signal to the input port.
11-6
Using Application Programs
= 10 log
P W R
10
Figure 11-5. 99 % Occupied Power Bandwidth
performs summation of the powers by
, and the frequency span,
X
x
=I2
(
)
[mW]
D
x
F span
N op
x
=I1
. The
(11 2)