Ecg Cable Protector Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-6

ECG Cable Protector Replacement Procedure

8-4-6-1
Tools
Appropriate Phillips screwdriver.
8-4-6-2
ECG Cable Protector Removal Procedure
1.) While facing the SafeLock Cart (i.e., working from the front of the Cart,) loosen and remove the two
screws that secure the ECG Cable Protector to the Top Shelf Assembly (refer to
page
2.) Remove the ECG Cable Protector from the Top Shelf Assembly.
3.)
4.)
.
8-4-6-3
ECG Cable Protector Installation Procedure
If this is a first-time installation of the ECG Cable Protector, it is necessary to remove the two plastic
guide holders from the Top Shelf Assembly, as shown in
(i.e., working from the front of the Cart,) these are located on the right side of the SafeLock Cart.
Plastic guide holders
Figure 8-78 Removing Plastic Guide Holders from Top Shelf Assembly
1.) Thoroughly clean the surface of the Top Shelf Assembly, in particular the recessed area from where
the ECG Cable Protector (or plastic guide holders) were removed.
2.) Position the ECG Cable Protector on the Top Shelf Assembly surface, ensuring that the recessed
groove on the protector aligns with the groove on the SafeLock Cart, as shown in
8-68
1
EVISION
8-62). See
Figure 8-78
on right hand side
Section 8-4 - SafeLock Cart Components Replacement
below.
V
V
IVID I N AND
IVID Q
Figure
8-78. While facing the SafeLock Cart
N S
M
ERVICE
ANUAL
Figure 8-72 on
Figure
8-79.