3-10-2-5
Stand-alone Vivid i n/ Vivid q N
No network connection needed.
3-10-2-6
"Sneaker Net" Environment
No network connection needed.
3-10-2-7
Connection from Vivid i n/ Vivid q N to an EchoPAC PC Workstation
Direct Cable Connection from Vivid i n/ Vivid q N to an EchoPAC PC Workstation
via a Crossover Cable
You will only need a Crossover Cable for network use to connect the two units this way.
1.) Connect one end of the Crossover Cable to the network connector on the Vivid i n/ Vivid q N.
2.) Connect the other end to the network connector to the EchoPAC PC Workstation, see Chapter 3 in
the EchoPAC PC Workstation Service Manual.
Connection via a Peer-to-Peer Network
You will need a network hub and one network cable for each unit connected to the hub.
See
3-10-2-1 "Ethernet Switch Connections" on page
Connection via Hospital Network
You will need one network cable to connect the Vivid i n/ Vivid q N to a wall outlet on the hospital's
network.
3-10-2-8
Connection between a Vivid i n/ Vivid q N and a DICOM Server on a Network
You will need one network cable.
1.) Connect one end of the cable to the Ethernet connector on the Vivid i n/ Vivid q N.
2.) Connect the other end of the cable to the wall outlet.
If a Peer-to-Peer Network is connected to the hospital's network, you may connect the
Vivid i n/ Vivid q N to the Peer-to-Peer Network.