GE Vivid i n Service Manual page 437

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
1
EVISION
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance.
Chapter 7 - Diagnostics/Troubleshooting
DC Offset Calibration on page 7 -
V
V
N S
IVID I N AND
IVID Q
25, and then repeat the test.
M
ERVICE
ANUAL
7-85

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid q nFl000110Fl000090Fq000001Vivid-iVivid-q

Table of Contents