As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Tx Voltages Test sequence is complete:
-
-
Note:
For this test, there is one report (Tx Voltages) which is listed in the Reports window on the
right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each fan are within the pre-defined Max and
Min voltage limits, as shown in
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
7-106
1
EVISION
Green: All recommended tests for these components have been completed and no problems
were identified.
Red:
Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
Figure 7-73 View Test Log - Tx Voltages Test
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
Section 7-5 - Monitoring Tests
Figure
7-73.
V
V
N S
IVID I N AND
IVID Q
Figure
7-73.
M
ERVICE
ANUAL