7) When finished reviewing the first report, trackball to the Exit button and press Set to close the View
Report window.
8) Select the second report (BPF1); the report is highlighted in the Reports list, and the corresponding
results are shown in the Report window below, as shown in
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the Close button and press Set to close the View Report window.
11) Select the third report (BPF2); the report is highlighted in the Reports list, and the corresponding
results are shown in the Report window below, as shown in
1
EVISION
Figure 7-49 View Test Log - ACW BPF Test - BPF1
Chapter 7 - Diagnostics/Troubleshooting
V
V
N S
IVID I N AND
IVID Q
Figure
7-49.
Figure
7-50.
M
ERVICE
ANUAL
7-77