Probe Shelf Assembly Replacement Procedure

NOTE:
The probe holders and rails are an integral part of the probe shelf assembly; these components may
not be replaced separately.
8-4-17-1
Tools
Appropriate Phillips screw driver.
8-4-17-2
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit - System Setup.
8-4-17-3
Probe Shelf Assembly Removal Procedure
1) Remove all peripherals from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Working from the front of the SafeLock Cart and underneath the probe shelf assembly, loosen and
remove the two lower screws (highlighted), that secure the shelf in position,
3) While supporting the shelf with one hand, loosen and remove the two upper screws (one on
either side - see
Upper Screws
8-90
1
EVISION
Figure 8-103 Probe Shelf Assembly - Lower Securing Screws
Figure
8-104); remove the shelf.
Figure 8-104 Probe Shelf Assembly - Upper Securing Screws