TR32 and RFI Boards Removal Procedure
1) Remove the bearing handle, as described in the
2) Remove the Control Panel and Keyboard, as described in the
Removal Procedure on page 8 -
3) Loosen the eight screws that secure the TR32 and RFI Boards to the Bottom Assembly, as shown
in
Figure 8-37
Figure 8-37 Loosen Screws that Secure TR32 and RFI Boards to Bottom Assembly
4) Remove the eight screws - see
5) Disconnect the PCI cable flex (BEP-to-RFI) from the cable connector on RFI Board as shown in
1
EVISION
below.
Figure 8-38 Removing the Screws
Chapter 8 - Replacement Procedures