Upper Cover (Rear) Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-3

Upper Cover (Rear) Replacement Procedure

8-4-3-1
Tools
Appropriate Phillips screwdriver.
8-4-3-2
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-3-3
Upper Cover (Rear) Removal Procedure
Remove the retaining screws of the upper rear cover and while supporting the cover with one hand,
carefully remove the cover (see
8-4-3-4
Upper Cover (Rear) Installation Procedure
1.) Return the cover to the correct position on the rear of the SafeLock Cart (see
2) While holding the cover to prevent it from falling, return and fasten the two upper screws
(see
3) Fasten the four remaining screws that secure the upper cover (rear).
4) Proceed to perform the following functionality tests:
-
-
8-64
1
EVISION
Figure 8-73 Upper Cover (Rear) Removal
Figure
8-73).
SafeLock Cart - Grounding Continuity on page 10 - 23
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Section 8-4 - SafeLock Cart Components Replacement
Figure
8-73).
V
V
N S
IVID I N AND
IVID Q
Retaining Screw
Figure
M
ERVICE
ANUAL
8-73).