LCD Display Adjustments and Calibration

Overall low image quality or "shades" visible on the Vivid i n/ Vivid q N LCD Display may be the result
of improper monitor calibration. Therefore, adjustments may be made to the Contrast, Brightness,
Watermark and/or Blue Tint, as required.
Important: LCD Display adjustments should be made in the correct order, as stated in the following
procedures.
NOTE:
Incorrect calibration of the Vivid i n/ Vivid q N monitor may also result in less than optimal image quality
when reviewing images on a DICOM or EchoPAC station.
Access the Calibration Screen (by pressing <ALT>+<Q>), then follow the relevant procedure
instructions described below.
NOTE:
On systems with older software versions, press <CTRL>+<ALT>+<P> to enter the Calibration Screen.
NOTE:
The Hot keys used to exit the Calibration Screen are: <ALT>+<Q>.
6-2-1

Contrast Adjustment Procedure

To adjust the Contrast: hold down <ALT>:
-
-
6-2
1
EVISION
Figure 6-1 LCD Display Calibration Screen
Repeatedly pressing Steer
Repeatedly pressing Steer
Section 6-2 - LCD Display Adjustments and Calibration
V
IVID I N AND
to decrease the Contrast;
to increase the Contrast.
V
N S
M
IVID Q
ERVICE
ANUAL