GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-15-4
Top Shelf Installation Procedure
1) Carefully position the Top Shelf on the upper assembly and replace and fasten the rear and lower
retaining screws.
2) Replace and reconnect the Main Harness, Peripheral power cable as described in
Assembly Installation Procedure on page 8 - 73
3) Install the Gas Spring as described in Gas Spring Removal Procedure section, page 8-82.
4) Return the USB Box as described in the
5) Fasten all cables in position using new cable ties.
6) Return the Lower Cover (Rear) as described in the
page 8 -
7) Return the Upper Cover (Rear) as described in the
page 8 -
8) Reconnect all required peripherals to the SafeLock Cart.
9) Reconnect the AC adapter and all cables to the Vivid i n/Vivid q N portable ultrasound scanner.
10) Proceed to perform the following functionality tests:
-
-
8-88
1
EVISION
Figure 8-101 Top Shelf - Rear Retaining Screws
71.
64.
SafeLock Cart - Grounding Continuity on page 10 - 23
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Section 8-4 - SafeLock Cart Components Replacement
V
IVID I N AND
and connect the LAN cable.
USB Board Installation Procedure on page 8 -
Lower Cover (Rear) Installation Procedure on
Upper Cover (Rear) Installation Procedure on
V
N S
M
IVID Q
ERVICE
ANUAL
Top Shelf rear retaining
screws
AC Distribution
80.