GE Vivid i n Service Manual page 83

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
3-5-3-1
Damage Inspection Checklist (cont'd)
Table 3-8
b
1
EVISION
Damage Inspection Checklist - SafeLock Cart (optional)
Step
Item
1
Probe Holders
Peripheral
2
Power and USB
Audio Panels
3
Covers
4
Peripherals
5
AC System
6
Power Cord
7
Front Castors
8
Rear Castors
Chapter 3 - System Setup
Recommended Procedure
Clean the gel wells with warm water and a damp cloth to remove all traces of gel.
Check the panel connectors for bent pins, loose connections and loose or missing
hardware. Screw all the cable connectors tightly to the connector sockets on the
panel. Verify that the labeling is in good condition.
Check that all screws are in place, all chassis and internal covers are installed.
Check and clean the peripherals in accordance with the manufacturer's directions.
To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.
Check the AC board connectors and the associated cabling for good connection
and proper insulation. Verify that the connections are secured.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or
any deterioration. Verify continuity.
Tighten the clamps that secure the power cord to the unit and the outlet plug to the
cord. Replace the power cord and/or clamp, as required.
Check that the front castors can swivel, and can be placed in the locked position
by pressing the foot brake lever (lower) down on each castor. Check that the upper
lever releases the SafeLock Cart wheel lock on each castor wheel.
Check that the rear castors can roll and swivel.
V
V
N S
IVID I N AND
IVID Q
Clamp securing power cord
M
ERVICE
ANUAL
3-15