Rfi Board - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
5-3-3

RFI Board

The RFI Board combines the following functionalities:
Front End Control
Radio Frequency & Tissue Processing
PCI Bus Interface
5-3-3-1
Tasks Performed by the RFI Board
System Master Clock generator
Transmitter Control
ATGC Control
Receive Focusing Control
Analog Test Signal Generator
Tx voltage Enable/Disable
Probe Sensing
Control of TR32 and P&M Boards
Ultrasound data acquisition and processing
5-3-3-2
Watchdog and Power ON Reset
For safety reasons, a HW watchdog stops ultrasound transmit and resets the DSP if the DSP stops
toggling the watchdog inputs.
Once enabled, it is not possible to disable the watchdog unless the board is reset, or power is cycled.
5-3-3-3
Boot-up of DSP and Firmware
When the power is turned ON, the DSP is reset, and all FPGAs are blank. The DSP is connected to PCI
directly, and can only boot up with help from the Host.
1
EVISION
Figure 5-8 RFI Block Diagram
Chapter 5 - Components and Function (Theory)
V
V
N S
IVID I N AND
IVID Q
M
ERVICE
ANUAL
5-11