GE Vivid i n Service Manual page 487

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
3.) Select and drag the Test screen to another region so that the display screen is not obscured.
4.) Trackball to the required button (Solid Color Test: Red, Green or Blue) and press Set to perform
the corresponding test.
5.) When in the displayed screen, press any key to return to PC Dr for Windows Monitor Test screen.
6.) When done, trackball to the Close button and press Set to exit the test.
When the Monitor Test sequence is complete:
-
-
1
EVISION
Figure 7-100 PC Doctor for Windows Monitor Test
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in
Figure 7-101 View Test Log - Monitor Test