IRECTION
8-5-3

BIOS Firmware Update

WARNING: When performing this procedure, it is strongly recommended that both AC power
and battery power be available to the Vivid i n/ Vivid q N scanner to prevent causing problems
in the Back End in the event of AC power fluctuation or power failure.
In addition, BIOS Firmware Update is a sensitive procedure; allow the process to run to
completion without interruption. Failure to do this could result in unrecoverable damage to the
system!
1) Make sure the Vivid i n/ Vivid q N scanner is connected to both battery and AC power.
2) Insert the Vivid i n/ Vivid q N software installation CD in the DVD\CD-R and close the tray.
3) Hold down the ON/OFF button on the control console for 3 seconds.
The system powers ON and when system boot-up from the CD is complete, the following options
are displayed in the DOS screen, prompting you to select the required option:
4) Select option: 3: BIOS Update MODBR949
A prompt for password is displayed:
5) Type geulsservice and press <Enter>.
The following screen opens displaying a flashing Warning message not to shutdown the system
during this procedure:
Note: use of the battery for DC backup during this procedure is also recommended in this screen.