GE Vivid i n Service Manual page 427

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
7-4-5-2-5
ACW BPF Test
NOTE:
For a description of this test, refer to the information in
IMPORTANT In order to access an individual board test, it is necessary to insert the service dongle from
NOTICE
system boot-up and enter the Service Password.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be touching
the probe connectors!
It is necessary to disable the Loop option where it should not be used.
After performing this test, you must re-boot the system and perform the DC Offset Calibration procedure
as described in
1.) In the Diagnostic Test window, trackball to ACW BPF Test in the list of ACW tests available for
selection from the Diagnostic Test Tree (see
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (P&M), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
1
EVISION
DC Offset Calibration on page 7 -
Figure 7-47 ACW BPF Test
Chapter 7 - Diagnostics/Troubleshooting
V
IVID I N AND
ACW BPF Test on page 7 -
25.
below) and press Set. The ACW BPF Test
Figure 7-47
V
N S
M
IVID Q
ERVICE
ANUAL
13.
7-75