15.)Use the tools provided to scroll through the channels and view the results, as previously described.
16) When done, trackball to the Close button and press Set to close the View Report window.
17) After reviewing the test results, proceed as follows:
18.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
19.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
1
EVISION
Figure 7-51 View Test Log - ACW BPF Test - BPF3
Re-boot the system (this is obligatory).
Perform calibration, as described in
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance.
Chapter 7 - Diagnostics/Troubleshooting
V
IVID I N AND
DC Offset Calibration on page 7 -
V
N S
M
IVID Q
ERVICE
ANUAL
25.
7-79