IMPORTANT: Safety Note
The VGA output of the Vivid i n/ Vivid q N is a non-isolated output, therefore it may only be connected
directly to a medical device.
In cases where the VGA contact is intended to be connected to a non-medical device, it is
recommended to implement one of the following solutions:
1.) Purchase a standard VGA isolated splitter. Do not use the GE Healthcare VGA isolated splitter
(P/N 5390022) as this has a special pin configuration (dedicated for use with CardioLab/MacLab/
Carto).
2.) Connect the device to an isolation transformer. Ensure that the device is not connected to another
non-medical device which can cause isolation failure through the third device.
3.) Power the second device via a medical grade power supply. Ensure that the device is not
connected to another non-medical device which can cause isolation failure through the third device.
1
EVISION
Chapter 6 - Service Adjustments
V
V
N S
IVID I N AND
IVID Q
M
ERVICE
ANUAL
6-9