GE H
EALTHCARE
D
FQ091013, R
IRECTION
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full Video Test sequence is complete:
-
-
Note:
For the Full Video Test, several tests are performed in the test sequence; these are listed
in the Tests window as shown in
Video Memory) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
-
-
-
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
1
EVISION
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in
Figure 7-91 View Test Log - Full Video Test
Video Memory
Video Data Transfer
Video Text Output
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance.
Chapter 7 - Diagnostics/Troubleshooting
Figure
7-91. For each currently-selected test (for example,
V
V
N S
IVID I N AND
IVID Q
Figure
M
ERVICE
ANUAL
7-91.
7-127