Peripherals - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
4-3-1-11
Visual Test
Table 4-15
Step
1
2
4-3-2

Peripherals

4-3-2-1
VCR Test
Table 4-16
Step
1
2
3
4
5
6
4-28
1
EVISION
Visual Test - Recommended
Task
Insert 3S-RS probe.
Increase the gain to see the homogenous noise.
VCR Test - Recommended
Task
Connect a Mitsubishi 9600 VCR and video converter to
the system, restart the system and load an SVHS
cassette.
Connect an external monitor to the VCR S-Video or
Video output.
Connect a 3S-RS probe. Reset the VCR counter and
press the VCR Record button.
Activate CW Mode and record 2 minutes of heart
scanning or a similar effect using a phantom.
Stop recording and rewind the cassette. Press the Play
button on the VCR.
Section 4-3 - Functional Checks
V
V
IVID I N AND
IVID Q
Expected Result(s)
The probe ID is 3S-RS. No strikes or artefacts are
present.
Expected Result(s)
Verify that you follow all the procedure instructions
described in
Connecting the VCR on page
Verify that the recording indicator (red dot) is displayed
on the VCR and the tape counter is running on the
VCR.
Record 2 minutes of CW scanning.
The recorded Image is played back.
Verify that in CW Mode both the left and right audio
channels are working, and that the played-back audio
is similar to the recorded audio in both the left and right
speakers, when applicable.
N S
M
ERVICE
ANUAL
3-48.