GE H
EALTHCARE
D
FQ091013, R
IRECTION
2-2-5-5
Power Stability Requirements
Voltage drop-out
Max 10 ms.
Note:
Only relevant if system is operating without a battery.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
2-2-6

EMI Limitations

Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid i n and Vivid q N
ultrasound units comply with limits as stated on the EMC label. However, there is no guarantee that
interference will not occur in a particular installation.
NOTE:
Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
Medical lasers.
Scanners.
Cauterizing guns.
Computers.
Monitors.
Fans.
Gel warmers.
Microwave ovens.
Portable phones.
Broadcast stations and mobile broadcasting machines.
2-4
1
EVISION
Section 2-2 - Console Requirements
V
V
N S
IVID I N AND
IVID Q
M
ERVICE
ANUAL