Lcd Display Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-3-8

LCD Display Replacement Procedure

8-3-8-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Display replacement
procedure.
8-3-8-2
Time Required
15 minutes
8-3-8-3
Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-8-4
LCD Display Removal Procedure
1) Remove the bearing handle, as described in the
2) Remove the Control Panel and Keyboard as described in the
Procedure on page 8 -
3.) Remove the LCD Display Frame as described in the
page 8 -
4.) Loosen and remove the four screws that fasten the LCD Display to the Rear Cover & Latch
Assembly, as shown in shown in
5.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in
Use a flat screwdriver to carefully release the cable from the connector.
1
EVISION
9.
27.
Figure 8-62 Removing the Screws from the LCD Display
Figure 8-63 Disconnecting the LCD Flex Ribbon Cable
Chapter 8 - Replacement Procedures
Bearing Handle Removal Procedure on page 8 -
LCD Display Frame Removal Procedure on
Figure 8-62
below.
V
V
N S
IVID I N AND
IVID Q
Control Panel and Keyboard Removal
Figure 8-63
below.