CAUTION: The battery has a safety device.
-
-
-
-
-
-
-
-
-
CAUTION: If the Vivid i n or Vivid q N portable ultrasound system is not being used for periods of one
month or longer, the battery should be removed during lengthy non-use period.
CAUTION: To avoid the battery bursting, igniting, or fumes from the battery causing equipment
damage, observe the following precautions:
-
-
-
-
Recommended storage conditions of battery pack:
Short term (less than one month):
Long term (more than three months): 10° C (50° F) to 35° C (95°F).
1
EVISION
Do not disassemble or alter it. Charge the batteries only when the ambient temperature is
between 0° and 65° C (32° and 149° F) and discharge the batteries between -10° and 55° C
(14° and 131° F).
Do not short-circuit the battery by directly connecting the battery terminals with metal objects.
Do not heat the battery or incinerate.
Do not expose the battery to temperature over 60° C (140° F). Keep it away from fire and
other heat sources.
Do not charge the battery near a heat source, e.g. fire or heaters.
Do not leave the battery in direct sunlight.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery. Do not solder a battery.
Do not connect the battery to an electrical outlet.
Do not immerse the battery in water or allow it to get wet.
Do not put the battery into a microwave oven or pressurized container. If the battery leaks or
emits an odor, remove it from all possible flammable sources.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it.
If you have any questions about the battery, consult GE or your local representative.