Battery Calibration - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
EVISION
6-2-11

Battery Calibration

NOTE:
It is recommended to perform the Battery Calibration procedure when the battery is fully
charged (as much as the system allows) and preferably to commence the procedure towards
the end of the working day - then allow it to continue to run automatically, overnight.
For details, see
IMPORTANT: During the Battery Calibration procedure, no scanning can be performed!
Towards end of day
Overnight
Next morning
Note:
For step-by-step instructions, see
12.
1
Figure 6-10
below.
F ULL CHARGE
F ULL DISCHARGE
Vivid i n/ Vivid q N System
Figure 6-10 Recommendations - Battery Calibration Procedure
Chapter 6 - Service Adjustments
ready for use
Performing the Battery Calibration Procedure on page 6 -
V
V
N S
IVID I N AND
IVID Q
May take up to 4 hours
May take up to 2 hours
Battery calibrated
M
ERVICE
ANUAL
6-11