Front Wheel Installation Procedure
1) Place the new front wheel on the SafeLock Cart chassis in the same position as that of the
previously removed wheel.
2) Return and fasten the four bolts previously removed and make sure that wheel is secured firmly to
the chassis.
3) Stand the SafeLock Cart upright.
4) Return the Vivid i n/ Vivid q N and all peripherals to their appropriate positions on the SafeLock Cart.
5) Re-connect the AC adapter and all cables and peripherals to the Vivid i n/Vivid q N portable
ultrasound scanner.
6) Re-connect mains power to the AC adapter.
7) Proceed to perform the following functionality tests:
-
-
8-94
1
EVISION
SafeLock Cart - Grounding Continuity on page 10 - 23
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25