GE Vivid i n Service Manual page 454

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
7-5-1-5
Battery Parameters Test
NOTE:
For a description of this test, refer to the information in
IMPORTANT In order to access an individual board test, it is necessary to insert the service dongle from
NOTICE
system boot-up and enter the Service Password.
1.) In the Diagnostic Test window, trackball to Battery Parameters in the list of Monitoring tests
available for selection from the Diagnostic Test Tree (see
Battery Parameters Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Battery Parameters Test commences. While the test proceeds, the name of the test currently
in progress (Battery Parameters) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
7-102
1
EVISION
Figure 7-70 Battery Parameters Test
Section 7-5 - Monitoring Tests
V
V
IVID I N AND
IVID Q
Battery Parameters Test on page 7 -
below) and press Set. The
Figure 7-70
N S
M
ERVICE
ANUAL
14.