TR32 and RFI Boards Replacement Procedure

NOTE:
When replacing the TR32 Boards, both boards must be of the same version. Always verify this by
examining the version number stated on the board, as shown in
CAUTION: When removing the boards, ESD may cause damage to a board. Always have the ESD
hand strip connected to the machine chassis and to your hand.
8-3-2-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-3-2-2
Time Required
15 minutes
8-3-2-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-32
1
EVISION
Figure 8-36 TR32 Board Verification (Version 12)
Section 8-3 - Internal Component Replacement Procedures
V
V
N S
IVID I N AND
IVID Q
Figure
8-36, below.
M
ERVICE
ANUAL
TR32 Board
Version (V12)
Verification