Test Setup
Power Meter
Figure 2-1 Zero Test Setup
Procedure
The following procedure should be performed for the Zero Test.
The procedure details the key presses required on the Agilent Technologies E4416A.
N O TE
For the Agilent Technologies E4417A, the equivalent key presses should be performed
on both channels.
1 Connect the equipment as shown in
2 Switch the power meter on.
3 Press
4 Press
5 Set the range calibrator as follows:
• RANGE: 3 μW
• POLARITY: NORMAL
• FUNCTION: STANDBY
• LINE: ON
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Range
Calibrator
CHANNEL A
Agilent 11730A
Preset
Confirm
then
.
Local
Meas
,
,
for a reading in watts.
Units W
Display
Performance Tests
Power
Meter
Figure 2-
1.
2
19