Performance Tests

Test Setup

Figure 2-9 Time Base Accuracy Test Setup

Procedure

Use the following procedure to perform the test:
1 Connect the equipment as shown in
2 Switch on the power meter.
3 Preset the instrument:
Press
4 Enable time base test mode:
Press
Time Base
5 Note the reading on the frequency counter and verify that it is within the
limits shown in
48
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power Meter (rear)
Trig out
Preset
then
Confirm
.
Local
System
More
,
,
Service
,
Enable 10 MHz TB
Table 2-
7:
Frequency Counter
Figure 2-
9.
More
,
,
,
Self Test
E4416A/E4417A Power Meters Service Guide

Hide quick links: