CONFigure Subsystem

CONFigure[:VOLTage]:AC

CONFigure[:VOLTage]:AC

Syntax
CONFigure[:VOLTage]:AC
This command first resets all AC voltage measurement parameters and trigger
parameters to their default values. Then, it configures the instrument for AC
voltage measurements but does not initiate the measurement.
C A U T I O N
N O T E
84
[<range>|AUTO|MAX|MIN|DEF[,{<resolution>|MAX|MIN|DEF}]]
• The maximum range parameter is 750 V
The rms voltage is waveform dependent. A sine wave is limited to
750 V
ac
AC MAINS are further limited to CAT II (300V).
• See the "Safety Information" section in the U3606B User's and
Service Guide for a complete discussion of the safety features, and
the procedures for safe operation of this instrument.
The
CONFigure[:VOLTage]:AC
instrument in the "wait-for-trigger" state. Use the
INITiate[:IMMediate]
CONFigure[:VOLTage]:AC
"wait-for-trigger" state.
(rms), but a 1000 V
square wave is safe. Connections to
pk
or
READ?
to place the instrument in the
, which is set by MAX.
rms
command does not place the
command in conjunction with the
U3606B Programmer's Reference