CALibration Subsystem
Remote Calibration Procedures
10 Select the 10 V range. Input 10 V, 1 kHz AC
11 Calibrate the 10 V point for AC voltage
12 Start the calibration.
13 Select the 100 V range. Input 100 V, 1 kHz
14 Calibrate the 100 V point for AC voltage
15 Start the calibration.
16 Select the 750 V range. Input 750 V, 1 kHz
17 Calibrate the 750 V point for AC voltage
18 Start the calibration.
62
Calibration steps
AC voltage gain adjustment — 10 V
voltage to the V (red) and LO (black) input
terminals.
measurements.
AC voltage gain adjustment — 100 V
AC voltage to the V (red) and LO (black)
input terminals.
measurements.
AC voltage gain adjustment — 750 V
AC voltage to the V (red) and LO (black)
input terminals.
measurements.
Remote commands
& CONF:VOLT:AC 10
& CAL:VAL 10
& CAL?
& CONF:VOLT:AC 100
& CAL:VAL 100
& CAL?
& CONF:VOLT:AC 750
& CAL:VAL 750
& CAL?
U3606B Programmer's Reference