OUTPut Subsystem

OUTPut[:STATe]

OUTPut[:STATe]
Syntax
OUTPut[:STATe] {0|1|OFF|ON}
This command enables or disables (output on standby) the source output of
the source multimeter.
The valid options are:
• 0|OFF: Output is on standby (the SBY annunciator turns on).
• 1|ON: Output is active (the OUT annunciator turns on).
N O T E
OUTPut[:STATe]?
This query returns a boolean value that represents the instrument output
state. A returned value of "1" indicates that the output is currently active. A
returned value of "0" indicates that the output is currently on standby.
Parameter
state
Example
& OUTP ON
& OUTP?
$ 1
168
The ramp and scan output will be restarted when output is enabled.
Item
Type
Boolean
Range of values
0|1|OFF|ON
This command enables the instrument output.
This query returns the output status of the
instrument.
Default value
0
U3606B Programmer's Reference