Agilent Technologies U3606B Programmer's Reference Manual page 69

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

7 Remove the shorting plug from the input
8 Calibrate the open point for 2-wire
9 Start the calibration.
10 Select the DC current measurement. Leave
11 Calibrate the open point for DC current
12 Start the calibration.
13 Select the capacitance measurement. Leave
14 Calibrate the open point for capacitance
15 Start the calibration.
U3606B Programmer's Reference
Calibration steps
Zero offset adjustment — 2-wire resistance (open)
terminals (all terminals open). Select the
2-wire resistance measurement, 100 MΩ
range.
resistance measurements.
Zero offset adjustment — DC current (open)
the input terminals open.
measurements.
Zero offset adjustment — Capacitance (open)
the input terminals open.
measurements.
CALibration Subsystem
Remote Calibration Procedures
Remote commands
& CONF:RES 100M
& CAL:VAL 9.9E+37
& CAL?
& CONF:CURR
& CAL:VAL 9.9E+37
& CAL?
& CONF:CAP
& CAL:VAL 9.9E+37
& CAL?
3
57