AC current gain adjustment procedure
Follow the steps outlined below. Review the "Test Considerations" described in
the U3606B Service Guide and the
page 58 before beginning this test.
1 Select the AC current measurement, 1 mA
2 Calibrate the 1 mA point for AC current
3 Start the calibration.
4 Select the 10 mA range. Input 10 mA, 1 kHz
5 Calibrate the 10 mA point for AC current
6 Start the calibration.
7 Select the 100 mA range. Input 100 mA,
8 Calibrate the 100 mA point for AC current
9 Start the calibration.
U3606B Programmer's Reference
Calibration steps
AC current gain adjustment — 1 mA
range. Input 1 mA, 1 kHz AC current to the
I (red) and LO (black) input terminals.
measurements.
AC current gain adjustment — 10 mA
AC current to the I (red) and LO (black) input
terminals.
measurements.
AC current gain adjustment — 100 mA
1 kHz AC current to the I (red) and LO (black)
input terminals.
measurements.
CALibration Subsystem
Remote Calibration Procedures
"Gain adjustment considerations"
Remote commands
& CONF:CURR:AC 0.001
& CAL:VAL 0.001
& CAL?
& CONF:CURR:AC 0.01
& CAL:VAL 0.01
& CAL?
& CONF:CURR:AC 0.1
& CAL:VAL 0.1
& CAL?
3