[SOURce:]CURRent[:LEVel]:SCAN[:AMPLitude]
Syntax
[SOURce:]CURRent[:LEVel]:SCAN[:AMPLitude] <value>
This command sets the amplitude end position for the current scan signal.
[SOURce:]CURRent[:LEVel]:SCAN[:AMPLitude]?
This query returns a numeric value that represents the amplitude end position
of the current scan signal.
Parameter
value
Remarks
• When the output range is changed (SOURce:VOLTage:RANGe or
• The scan amplitude end position value is stored in the nonvolatile memory.
Example
& CURR:SCAN 1
& CURR:SCAN?
$ +1.000000E+00
U3606B Programmer's Reference
Item
Type
Numeric
SOURce:CURRent:RANGe command), the scan amplitude end position will
always be set to the default value (if unchanged) or set to the last stored
value.
It is not affected by a power-on cycle, Factory Reset
(*RST command), Instrument Preset (SYSTem:PRESet command), or
function change.
[SOURce:]CURRent[:LEVel]:SCAN[:AMPLitude]
Range of values
Range
Min
S1
0 A
S1m
0 A
S2
0 A
This command sets the amplitude end position
of the current scan signal to 1 A.
This query returns the amplitude end position
of the current scan signal.
SOURce Subsystem
Default value
Max
1.05 A
1 A
105 mA
100 mA
3.15 A
3 A
11
279