U3606B Programmer's Reference
Calibration steps
Resistance gain adjustment — 1 MΩ
resistance to the Ω (red) and LO (black)
input terminals.
resistance measurements.
Resistance gain adjustment — 100 kΩ
resistance to the Ω (red) and LO (black)
input terminals.
resistance measurements.
Resistance gain adjustment — 10 kΩ
resistance to the Ω (red) and LO (black)
input terminals.
resistance measurements.
Resistance gain adjustment — 1000 Ω
resistance to the Ω (red) and LO (black)
input terminals.
resistance measurements.
CALibration Subsystem
Remote Calibration Procedures
Remote commands
& CONF:RES 1M
& CAL:VAL 1M
& CAL?
& CONF:RES 100k
& CAL:VAL 100k
& CAL?
& CONF:RES 10k
& CAL:VAL 10k
& CAL?
& CONF:RES 1000
& CAL:VAL 1000
& CAL?