CALibration Subsystem
Remote Calibration Procedures
10 Select the 1 A range. Input 1 A DC current to
11 Calibrate the 1 A point for DC current
12 Start the calibration.
68
Calibration steps
DC current gain adjustment — 1 A
the I (red) and LO (black) input terminals.
measurements.
Remote commands
& CONF:CURR:DC 1
& CAL:VAL 1
& CAL?
U3606B Programmer's Reference