Calibration:count - Agilent Technologies U3606B Programmer's Reference Manual

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

CALibration:COUNt?

Syntax

CALibration:COUNt?

This query returns a numeric value that represents the calibration count
indicating how many calibrations have been performed in the instrument.
Note that your instrument was calibrated before it left the factory. When you
receive your instrument, be sure to read the count to determine the initial
values.
Remarks
• The calibration counts increment up to a maximum of 32767 after which
• The calibration count is incremented by the CALibration[:ALL]?
• The calibration count is stored in nonvolatile memory, and does not change
Example
& CAL:COUN?
$ +739
See also
"CALibration[:ALL]?"
"CALibration:SECure:CODE"
U3606B Programmer's Reference
they roll over to "0". Since the value increments by one for each calibration
point, a complete calibration may increase the value by many counts.
command. You can read the calibration count whether the instrument is
secured or unsecured.
when power has been off or after a Factory Reset (*RST command).
on page 46
This command returns the calibration count.
on page 48
CALibration Subsystem
CALibration:COUNt?