MEASure[:VOLTage]:ACDC|DCAC?

Syntax

MEASure[:VOLTage]:ACDC|DCAC?

This command first resets all AC+DC voltage measurement parameters and
trigger parameters to their default values. Then, it configures the instrument
for AC+DC voltage measurements and immediately triggers a measurement. A
single reading is returned.
C A U T I O N
Parameters
range
resolution
U3606B Programmer's Reference
[<range>|AUTO|MAX|MIN|DEF[,{<resolution>|MAX|MIN|DEF}]]
• The maximum range parameter is 750 V
The rms voltage is waveform dependent. A sine wave is limited to
750 V
AC MAINS are further limited to CAT II (300V).
• See the "Safety Information" section in the U3606B User's and
Service Guide for a complete discussion of the safety features, and
the procedures for safe operation of this instrument.