Agilent Technologies U3606B Programmer's Reference Manual page 72

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

3
CALibration Subsystem
Remote Calibration Procedures
10 Calibrate the –1 V point for DC voltage
11 Start the calibration.
12 Select the 10 V range. Input 10 V DC voltage
13 Calibrate the 10 V point for DC voltage
14 Start the calibration.
15 Select the 100 V range. Input 100 V DC
16 Calibrate the 100 V point for DC voltage
17 Start the calibration.
18 Select the 1000 V range. Input 1000 V DC
19 Calibrate the 1000 V point for DC voltage
20 Start the calibration.
60
Calibration steps
measurements.
DC voltage gain adjustment — 10 V
to the V (red) and LO (black) input terminals.
measurements.
DC voltage gain adjustment — 100 V
voltage to the V (red) and LO (black) input
terminals.
measurements.
DC voltage gain adjustment — 1000 V
voltage to the V (red) and LO (black) input
terminals.
measurements.
Remote commands
& CAL:VAL -1
& CAL?
& CONF:VOLT:DC 10
& CAL:VAL 10
& CAL?
& CONF:VOLT:DC 100
& CAL:VAL 100
& CAL?
& CONF:VOLT:DC 1000
& CAL:VAL 1000
& CAL?