Frequency gain adjustment procedure
Follow the steps outlined below. Review the "Test Considerations" described in
the U3606B Service Guide and the
page 58 before beginning this test.
1 Select the AC voltage measurement, 1 V
2 Select the frequency measurement. Input
3 Calibrate the 1 kHz point for frequency
4 Start the calibration.
U3606B Programmer's Reference
Calibration steps
Frequency gain adjustment — 1 kHz
range.
1 V, 1 kHz AC voltage to the V (red) and
LO (black) input terminals.
measurements.
CALibration Subsystem
Remote Calibration Procedures
"Gain adjustment considerations"
Remote commands
& CONF:VOLT:AC 1
& CONF:FREQ
& CAL:VAL 1000
& CAL?