16
IEEE-488.2 Common Commands

*SAV

*SAV
Syntax
*SAV {0|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16}
The *SAV (SAVe) command stores (saves) the current instrument state in the
designated storage location. Any state previously stored in the same location
is overwritten (with no error generated).
Use the *RCL command to recall a stored instrument state.
Parameter
state
Remarks
• The instrument has sixteen storage locations in nonvolatile memory to
• When shipped from the factory, storage locations 1 through 16 are empty
• A Factory Reset (*RST command) or SYSTem:PRESet command does not
Example
& *SAV 1
See also
"*RCL"
346
Item
Type
Discrete
store instrument states. The instrument uses location 0 to automatically
hold the state of the instrument at power down. You can only recall a state
from a location that contains a previously stored state.
and the instrument is configured to automatically recall the power-down
state (location 0) when power is restored. You can change the factory
configuration such that a Factory Reset (*RST command) is issued when
power is restored.
affect the configurations stored in memory. Once a state is stored, it
remains until it is overwritten.
on page 341
Range of values
0 to 16
This command stores the current instrument
state in location 1.
Default value
-
U3606B Programmer's Reference