Remarks
• After selecting the trigger source, you must place the instrument in the
N O T E
• The CONFigure commands and MEASure? queries automatically sets the
• For the IMMediate (continuous) source, the trigger signal is always
• For the BUS (software) source, the instrument is triggered by the *TRG
• The instrument selects the IMMediate trigger source after a Factory Reset
• If the selected trigger source is invalid due to a settings conflict, the source
U3606B Programmer's Reference
"wait-for-trigger" state by using the INITiate[:IMMediate] command or
READ? query. A trigger will not be accepted from the selected trigger source
until the instrument is in the "wait-for-trigger" state.
The INITiate[:IMMediate] command only initiates the
measurement and needs a trigger (BUS or IMMediate) to make the
actual measurement.
trigger source to IMMediate.
present. When you place the instrument in the "wait-for-trigger" state, the
trigger is issued immediately.
command received over the remote interface. The *TRG command will not
be accepted unless the internal instrument is in the "wait-for-trigger" state.
(*RST command) or an Instrument Preset (SYSTem:PRESet command).
reverts automatically to IMMediate. A settings conflict error message is
also generated.
TRIGger Subsystem
TRIGger:SOURce
325