16
IEEE-488.2 Common Commands
*TRG
*TRG
Syntax
*TRG
The *TRG (TRiGger) is used in conjunction with the TRIGger:SOURCe
command to trigger the instrument from the remote interface.
Remarks
• Use the TRIGger:SOURce command to select the BUS (software) trigger
• After setting the trigger source, you must place the multimeter in the
Example
& TRIG:SOUR BUS
& INIT
& *TRG
See also
"INITiate[:IMMediate]"
"TRIGger:SOURce"
350
source.
"wait-for-trigger" state using the INITiate[:IMMediate] command. The
*TRG command will not be accepted unless the multimeter is in the
"wait-for-trigger" state.
on page 324
This command selects the bus (software)
trigger source.
This command places the instrument in the
"wait-for-trigger" state.
This command triggers the instrument.
on page 120
U3606B Programmer's Reference